Sign In

Primary Care Emergency Medicine, Resuscitation Stretcher Trolleys
Expanded Image

Get Quotation

Add To Favorites

Compare This Product

MRI stretcher trolley
AWL-3065/VA
AGA Sanitätsartikel
Characteristics
accessories with IV pole, with side rails
applications MRI, transport
features with adjustable backrest
number of sections 2-section
operation manual
other characteristics non-magnetic, stainless steel
Length 1900 mm
Maximum load 200 kg
wheel diameter 100 mm
Width 650 mm, 795 mm, 800 mm, 975 mm
Description
The MRI tables with side guards from AGA Sanitätsartikel were specially designed for use in magnetic resonance (abbreviated MR or MRI). The frame components are made of slightly magnetic stainless steel or aluminium and brass. The low level of residual magnetism does not affect the imaging process, nor are the MRI patient transport tables attracted by the magnetic field of the MR system. It can be used on MR systems up to 7 Tesla without any problems.

The two side guards included in the scope of delivery serve as fall protection and for safer patient transport.

These MRI patient transport tables can be used safely under the following conditions:

Static magnetic field of 7 Tesla or less,
With an additional spatial gradient of maximum 7 T/m or less,
The table must not exceed the limit line of 7 T/m with the head end/foot end in the direction of movement towards the magnet.

Features

The MRI patient transport couches are equipped with swivel castors, Ø 100 mm. All four can be totally locked in turning and running directions.
Two side guards are supplied with these MRI tables for safer patient transport.
The side guards, dimensions approx. 800 x 250 mm (L x H), can be lowered to the upholstery height using a pivot mechanism and secured using spring-loaded handwheels with fixing bolts.
The stability and robustness is achieved by the gusset plates in the frame corners and the cross and longitudinal braces between the legs.
The tubular steel frame of the MRI tables is made of stainless steel. It has a low residual magnetism.
It can be used with a static magnetic field of a maximum of 7 Tesla or less.

AGA Sanitätsartikel

0 Followers
  Löhner Strasse 198b 32584 Löhne
  73182283
  https://www.agasan.com
Products

Buyers Who Liked This Product Also Liked

Recently Viewed Products

Support cushion
TS-N63-4 series